Focus on draeger vulnerabilities and metrics.
Last updated: 01 Oct 2026, 22:25 UTC
This page consolidates all known Common Vulnerabilities and Exposures (CVEs) associated with draeger. We track both calendar-based metrics (using fixed periods) and rolling metrics (using gliding windows) to give you a comprehensive view of security trends and risk evolution. Use these insights to assess risk and plan your patching strategy.
For a broader perspective on cybersecurity threats, explore the comprehensive list of CVEs by vendor and product. Stay updated on critical vulnerabilities affecting major software and hardware providers.
Total draeger CVEs: 8
Earliest CVE date: 28 Jan 2019, 21:29 UTC
Latest CVE date: 02 Jun 2026, 14:16 UTC
Latest CVE reference: CVE-2019-25717
30-day Count (Rolling): 0
365-day Count (Rolling): 3
Calendar-based Variation
Calendar-based Variation compares a fixed calendar period (e.g., this month versus the same month last year), while Rolling Growth Rate uses a continuous window (e.g., last 30 days versus the previous 30 days) to capture trends independent of calendar boundaries.
Month Variation (Calendar): 0%
Year Variation (Calendar): 0%
Month Growth Rate (30-day Rolling): 0.0%
Year Growth Rate (365-day Rolling): 0.0%
Average CVSS: 3.35
Max CVSS: 7.2
Critical CVEs (≥9): 0
| Range | Count |
|---|---|
| 0.0-3.9 | 5 |
| 4.0-6.9 | 2 |
| 7.0-8.9 | 1 |
| 9.0-10.0 | 0 |
These are the five CVEs with the highest CVSS scores for draeger, sorted by severity first and recency.
Dräger Infinity Delta, Delta XL, and Kappa patient monitors contain an information disclosure vulnerability that allows unauthenticated network attackers to access log files over a network connection. Attackers can retrieve device internals, location information, and wired network configuration details from the exposed log files.
Dräger Infinity Explorer C700 contains a privilege escalation vulnerability that allows attackers to break out of kiosk mode and access the underlying operating system through a specific dialog interaction. Attackers can exploit this kiosk escape to take control of the operating system and cause the device to display incorrect or no information from the connected Delta Family patient monitor.
Dräger Infinity Delta, Delta XL, and Kappa patient monitors contain a denial-of-service vulnerability that allows remote attackers to cause the monitor to reboot by sending a malformed network packet. Attackers can repeatedly send malformed network packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.
Draeger X-Dock Firmware before 03.00.13 has Active Debug Code on a debug port, leading to remote code execution by an authenticated attacker.
Draeger X-Dock Firmware before 03.00.13 has Hard-Coded Credentials, leading to remote code execution by an authenticated attacker.
Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Log files are accessible over an unauthenticated network connection. By accessing the log files, an attacker is able to gain insights about internals of the patient monitor, the location of the monitor, and wired network configuration.
Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity.
Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Via a specific dialog it is possible to break out of the kiosk mode and reach the underlying operating system. By breaking out of the kiosk mode, an attacker is able to take control of the operating system.